SUIT-007: Difference between revisions
From Bioblast
No edit summary |
Garcia Luiz (talk | contribs) No edit summary |
||
Line 26: | Line 26: | ||
== Strengths and limitations == | == Strengths and limitations == | ||
:::* For determination of respirometric viability index, it is required cells with functional mitochondria from both intact and permeabilized (dead) cell population. | |||
:::+ S-ET capacity can be assessed after the performance of the [[SUIT-003]]. ce6S, 1Dig and 1c steps can be implemented without interfering other steps. | |||
:::+ Respirometric viability test correlates with other viability methods. | |||
:::+ CIV activity is measured. | |||
:::- Cells with succinate transporter are not suitable for the viability test. | |||
:::- Pathological profiles could hinder the viability test. | |||
:::- Longer protocol in comparison to [[SUIT-003]]. Protocol average length of 2 hours. | |||
:::- Careful washing is required after the experiment to avoid carry-over of inhibitors and uncoupler. | |||
== Compare SUIT protocols == | == Compare SUIT protocols == |
Revision as of 17:35, 15 January 2019
Description
Reference: A: ยปVersions
DLP applications
References
Steps and respiratory states
Step | State | Pathway | Q-junction | Comment - Events (E) and Marks (M) |
---|---|---|---|---|
1G | GL | (N) | CI | 1G
|
2D | GP | (N) | CI | 1G;2D
|
2c | GcP | (N) | CI | 1G;2D;2c
|
3M | GMP | N | CI | 1G;2D;2c;3M
|
4U | GME | N | CI | 1G;2D;2c;3M;4U
|
5Ama | ROX | 1G;2D;2c;3M;4U;5Ama
|
- Bioblast links: SUIT protocols - >>>>>>> - Click on [Expand] or [Collapse] - >>>>>>>
- Coupling control
- Pathway control
- ยป Electron transfer pathway
- ยป Fatty acid oxidation pathway control state, F
- ยป NADH electron transfer-pathway state, N
- ยป Succinate pathway control state, S
- ยป NS-pathway control state, NS
- ยป Glycerophosphate pathway control state, Gp
- ยป Complex IV single step, CIV
- ยป Anaplerotic pathway control state
- Pathway control
- Main fuel substrates
- ยป Glutamate, G
- ยป Glycerophosphate, Gp
- ยป Malate, M
- ยป Octanoylcarnitine, Oct
- ยป Pyruvate, P
- ยป Succinate, S
- Main fuel substrates
- Glossary
Strengths and limitations
- For determination of respirometric viability index, it is required cells with functional mitochondria from both intact and permeabilized (dead) cell population.
- + S-ET capacity can be assessed after the performance of the SUIT-003. ce6S, 1Dig and 1c steps can be implemented without interfering other steps.
- + Respirometric viability test correlates with other viability methods.
- + CIV activity is measured.
- - Cells with succinate transporter are not suitable for the viability test.
- - Pathological profiles could hinder the viability test.
- - Longer protocol in comparison to SUIT-003. Protocol average length of 2 hours.
- - Careful washing is required after the experiment to avoid carry-over of inhibitors and uncoupler.
Compare SUIT protocols
MitoPedia concepts:
MiP concept,
SUIT protocol,
Recommended
MitoPedia methods:
Respirometry