Talk:O2k-Open Support alert
From Bioblast
- Do you perform the leaky chamber test after chamber assembly?
- After chamber reassembly you might experience a small leaky chamber only identified by a leaky chamber test- Β»Leaky chamber testΒ«
- Do you perform the leaky chamber test after chamber assembly?
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - Β»Slope smoothingΒ«
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with specific settings for the O2k-sV-Module. - Β»O2k-sV-ModuleΒ«
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?