|description='''Volume calibration ring PVDF M2.5''' (A or B) PVDF for O2k- | |description='''Volume calibration ring PVDF M2.5''' (A or B) PVDF for O2k-Stopper sV, white; attached to the [[Stopper\black PEEK\conical Shaft\central Port |stoppers]]. | ||
|product id=2432A-01 or 2432B-01 (PRODUCT ID FOR A AND B?) | |product id=2432A-01 or 2432B-01 (PRODUCT ID FOR A AND B?) | ||
|product type=O2k, Stopper (TYPE STOPPER?) | |product type=O2k, Stopper (TYPE STOPPER?) |
Revision as of 16:48, 17 June 2019
User talk:Zhang Feiyuan/My sandbox 7
Description | Volume calibration ring PVDF M2.5 (A or B) PVDF for O2k-Stopper sV, white; attached to the stoppers. |
---|---|
Product ID | 2432A-01 or 2432B-01 (PRODUCT ID FOR A AND B?) |
Type | O2k, Stopper (TYPE STOPPER?) |
Link | MiPNet22.11 O2k-FluoRespirometer manual |
Image | TO REPACE |
CHECK THE TEXT BELOW
Technical support
- Volume calibration of the O2k-sV-Chamber
- Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
- Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
- Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
- Volume calibration of the O2k-sV-Chamber
Innovation
- Replacing the (white) PVDF Volume-Calibration Ring with the (black) aluminum Volume-Calibration Ring allows for more force applied to tighten the ring, thus reducing the risk of a loose fit and the associated danger of altering the calibrated position.