Template:Chemical SUIT-034 NADH mt D082: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 17: Line 17:
| 4U
| 4U
| {{Template:SF6847}}
| {{Template:SF6847}}
| We do not recommend the use of any other uncoupler, like {{Template:CCCP}}, due to background fluorescence.
| We do not recommend the use of any other uncoupler, like {{Template:CCCP}}, due to the chemical background effect on fluorescence.
|-
|-
| 5Anox
| 5Anox

Revision as of 16:28, 14 December 2023

Step Chemical(s) and link(s) Comments
1D.1 ADP (D)
2PGM Pyruvate (P), Glutamate (G), and Malate (M)
3D2.5 ADP (D)
4U SF6847 We do not recommend the use of any other uncoupler, like Carbonyl cyanide m-chlorophenyl hydrazone, CCCP (U), due to the chemical background effect on fluorescence.
5Anox The O2 concentration in the O2k-chamber can be decreased by N2 or H2 injection to reach faster anoxia, see: Setting the oxygen concentration.
6Myx Myxothiazol We do not recommend the use of any other inhibitor of complex III, like Antimycin A (Ama), due to the chemical background effect on fluorescence.
7Reox