'''Problem:''' After removing [[OroboPOS-Connector]], [[OroboPOS-Holder]], and [[O2k-Chamber Holder]]
from the O2k , the [[O2k-Chamber]]  cannot be removed. It remains stuck even after washing with water.
'''Solution''': The reason for this problem is very likely medium that has been spilled and acts now as a glue between glass chamber and copper block. 
Wash with plenty of warm water and try to remove the chamber while the O2k is switched on and heated up to 37 °C. You can try GENTLY to use the sensor holder (not the sensor connector) to try to push up the glass chamber (like you do during [[chamber assembly]]) - while the O2k is warm and with plenty of warm water.  Since the sensors are removed you also can use e.g. 40° (or a little bit more) for the washing / removing attempts.