== Switch componentsΒ  ==
Β 
{| cellspacing="1" cellpadding="1" border="1" align="center" width="513" style=""
|+ '''Summary'''
|-
! scope="col" | Problem moves with
! scope="col" | Problem situated at
|-
| [[POS]]
| [[POS]]
|-
| [[POS connector]]
| [[POS connector]]
|-
| not at all
| [[Chamber assembly]] or [[Main unit]]
|}
Β 
<br>
Β 
== Remove componentsΒ  ==
{| cellspacing="1" cellpadding="1" border="1" style="width: 745px; height: 114px;"
|+ '''Tested states and expected behavior'''
|-
| '''State'''
| '''Description'''
| '''Expected behavior'''
|-
| '''I'''
| fully assembled, sensor in chamber, air saturation, Gain =2
| stable signal, 3 to 5 V
|-
| '''II'''
| POS connector + sensor removed from chamber, sensor still connected to sensor connector
| initial change in voltage, new, stable non-zero level of signal reached
|-
| '''III'''
| sensor removed from POS connector
| stable signal at 0 V
|-
| '''IV'''
| POS connector removed from main unit
| for oxygraphs series A to C: typically signal off scale (+/- 9.99 V), for oxygraphs series D and higher: typically 0 V; may vary, a gain independent stable signal of 0.4 V is a strong indication of an electronic problem in the main unit
|}
Β 
<br>
Β 
{| cellspacing="1" cellpadding="1" border="1" style="width: 749px; height: 201px;"
|-
! scope="col" colspan="4" | State<br>expected behavior observed? yes/no <br>
! scope="col" rowspan="2" | Interpretation / Comment <br> <br>
|-
! scope="col" | I
! scope="col" | II
! scope="col" | III
! scope="col" | IV
|-
| no
| yes
| yes
| yes
| may indicate a problem in the [[Chamber assembly]], or a defective sensor, continue testing
|-
| no
| no/yes
| yes
| yes
| probable defect in the [[Sensor]]. This indication is not so strong as a localization obtained via switching components. The [[POS connector]] (not the [[Sensor]]) may be defect, but this may be only visible when a signal is present.
|-
| no
| no
| no
| yes
| indicates a defect in the [[POS connector]]
|-
| no
| no
| no
| no
| indicates a problem with [[Chamber assembly]] or a defect in the [[Main unit]]; if the problem can not be solved by reassembling the chamber, contact OROBOROS Instruments.
|}
Β 
<br>
{{#set:Technical service=O2 sensor|Technical service=POS connector| Technical service=Main unit}}