| {{Technical support}}
| | #REDIRECT [[O2k-Open Support]] |
| Β | |
| __TOC__
| |
| Β | |
| == Switch componentsΒ ==
| |
| Β | |
| {| cellspacing="1" cellpadding="1" border="1" align="center" width="513" style=""
| |
| |+ '''Summary'''
| |
| |-
| |
| ! scope="col" | Problem moves with
| |
| ! scope="col" | Problem situated at
| |
| |-
| |
| | [[POS]]
| |
| | [[POS]]
| |
| |-
| |
| | [[POS connector]]
| |
| | [[POS connector]]
| |
| |-
| |
| | not at all
| |
| | [[Chamber assembly]] or [[Main unit]]
| |
| |}
| |
| Β | |
| <br>
| |
| Β | |
| == Remove componentsΒ ==
| |
| {| cellspacing="1" cellpadding="1" border="1" style="width: 745px; height: 114px;"
| |
| |+ '''Tested states and expected behavior'''
| |
| |-
| |
| | '''State'''
| |
| | '''Description'''
| |
| | '''Expected behavior'''
| |
| |-
| |
| | '''I'''
| |
| | fully assembled, sensor in chamber, air saturation, Gain =2
| |
| | stable signal, 3 to 5 V
| |
| |-
| |
| | '''II'''
| |
| | POS connector + sensor removed from chamber, sensor still connected to sensor connector
| |
| | initial change in voltage, new, stable non-zero level of signal reached
| |
| |-
| |
| | '''III'''
| |
| | sensor removed from POS connector
| |
| | stable signal at 0 V
| |
| |-
| |
| | '''IV'''
| |
| | POS connector removed from main unit
| |
| | for oxygraphs series A to C: typically signal off scale (+/- 9.99 V), for oxygraphs series D and higher: typically 0 V; may vary, a gain independent stable signal of 0.4 V is a strong indication of an electronic problem in the main unit
| |
| |}
| |
| Β | |
| <br>
| |
| Β | |
| {| cellspacing="1" cellpadding="1" border="1" style="width: 749px; height: 201px;"
| |
| |-
| |
| ! scope="col" colspan="4" | State<br>expected behavior observed? yes/no <br>
| |
| ! scope="col" rowspan="2" | Interpretation / Comment <br> <br>
| |
| |-
| |
| ! scope="col" | I
| |
| ! scope="col" | II
| |
| ! scope="col" | III
| |
| ! scope="col" | IV
| |
| |-
| |
| | no
| |
| | yes
| |
| | yes
| |
| | yes
| |
| | may indicate a problem in the [[Chamber assembly]], or a defective sensor, continue testing
| |
| |-
| |
| | no
| |
| | no/yes
| |
| | yes
| |
| | yes
| |
| | probable defect in the [[Sensor]]. This indication is not so strong as a localization obtained via switching components. The [[POS connector]] (not the [[Sensor]]) may be defect, but this may be only visible when a signal is present.
| |
| |-
| |
| | no
| |
| | no
| |
| | no
| |
| | yes
| |
| | indicates a defect in the [[POS connector]]
| |
| |-
| |
| | no
| |
| | no
| |
| | no
| |
| | no
| |
| | indicates a problem with [[Chamber assembly]] or a defect in the [[Main unit]]; if the problem can not be solved by reassembling the chamber, contact OROBOROS Instruments.
| |
| |}
| |
| Β | |
| <br>
| |
| {{#set:Technical service=O2 sensor|Technical service=POS connector| Technical service=Main unit}}
| |