MiPNet15.05 NO-manual

From Bioblast
Publications in the MiPMap
Fasching M, Gnaiger E (2010) O2k-MultiSensor System with amperometric electrodes (NO, H2S, H2O2). Mitochondr Physiol Network 15.05.

Β» Open Access-MiPNet15.05; Versions

O2k-Manual contents

OROBOROS (2010) Mitochondr Physiol Network

Abstract: O2k-Manual Chapter I

>> O2k-Manual: Overall contents
>> Product: OROBOROS Oxygraph-2k, O2k-Catalogue

β€’ Keywords: HRR, MultiSensor

β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, TIP2k, NO, O2k-Manual 

O2k-SOP 


Corrigendum

The polarization voltages for the H2O2 sensors from WPI is model dependent. For the 2 mm sensor that usually will be used together with the O2k, the WPI recommendation is 800 to 900 mV, therefore -800 to -900 mV should be set in the O2k, not -400 mV as stated in MiPNET15.05. Fasching Mario 11:09, 20 January 2014 (CET)

Further information

>> O2k-NO Amp-Module
  • Version 3: 2011-12-11
  • Version 1: 2010