Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet15.08 TPP electrode"

From Bioblast
Line 1: Line 1:
{{Publication
{{Publication
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] Ion selective electrode for TPP<sup>+</sup> and high-resolution respirometry. [[Media:MiPNet15.08 TPP-electrode.pdf|»Bioblast pdf«]]
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] Ion selective electrode for TPP<sup>+</sup> and high-resolution respirometry.  
|info=[http://www.bioblast.at/index.php/File:MiPNet15.08_TPP-electrode.pdf Versions]
|info=[[File:PDF.jpg|50px|link=http://wiki.oroboros.at/images/c/c5/MiPNet15.08_TPP-electrode.pdf |Bioblast pdf]] »[http://www.bioblast.at/index.php/File:MiPNet15.08_TPP-electrode.pdf Versions]
|authors=OROBOROS
|authors=OROBOROS
|year=2011-12-11
|year=2011-12-11

Revision as of 16:51, 17 February 2016

Publications in the MiPMap
O2k-Protocols
Ion selective electrode for TPP+ and high-resolution respirometry.

» Bioblast pdf »Versions

OROBOROS (2011-12-11) Mitochondr Physiol Network

Abstract: Sumbalova Z, Fasching M, Gnaiger E (2011) Ion selective electrode for TPP+ and high-resolution respirometry. Mitochondr Physiol Network 15.8.

Tetraphenylphosphonium (TPP+) accumulates in the mitochondrial matrix as a function of the mitochondrial membrane potential. The TPP+ electrode is an ion selective electrode (ISE). The voltage signal [V] is linearly dependent on the logarithm of the free concentration [TPP+].

» Poster
» O2k-Protocols
» Product: Oxygraph-2k, ISE, O2k-Catalogue

Keywords: O2k-MultiSensor System

O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, TPP, O2k-Protocol 

O2k-Demo, O2k-MultiSensor