Oxygen signal not responding: Difference between revisions

From Bioblast
No edit summary
(Redirected page to Oxygen signal)
 
Line 1: Line 1:
{{Technical support}}
#REDIRECT [[Oxygen signal]]
 
 
== Problems ==
*The oxygen signal remains at or close to zero even at high oxygen levels.
*The signal at air saturation is abnormally low.
 
== Test and solutions ==
 
'''The oxygen signal is very low at air saturation and less responsive than expected.'''
:'''Solution 1:''' Check the [[gain]] setting for the affected oxygen channel.
::::*The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::*The gain was correct: Locate the problem using the procedure described in [[Locating a problem]] and [[Sensor test]].
 
 
{{#set: Technical service=O2 sensor|Technical service=O2 signal}}
 
 
__SHOWFACTBOX__
 
[[Category:Technical service]]