Oxygen signal not responding: Difference between revisions

From Bioblast
(remove obsolte part and send to "localizing a problem")
mNo edit summary
Line 6: Line 6:


'''The oxygen signal is very low at air saturation and less responsive than expected.'''  
'''The oxygen signal is very low at air saturation and less responsive than expected.'''  
:'''Solution 1:''' Check the gain setting for the oxygen.  
:'''Solution 1:''' Check the [[gain]] setting for the affected oxygen channel.
::::*The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::*The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
::::*The gain was correct: Locate the problem using the procedure described in [[Locating a Problem]] and [[Sensor test]]
::::*The gain was correct: Locate the problem using the procedure described in [[Locating a Problem]] and [[Sensor test]]




 
{{#set: Technical service= Sensor|Technical service= O2 signal}}
 
{{#set: Technical service= Sensor|Technical service= Signal}}


{{Technical service}}
{{Technical service}}


__SHOWFACTBOX__
__SHOWFACTBOX__
{{Troubleshooting}}

Revision as of 16:39, 13 September 2010

Problems

  • The oxygen signal remains at or close to zero even at high oxygen levels.
  • The signal at air saturation is abnormally low.

Test and Solutions

The oxygen signal is very low at air saturation and less responsive than expected.

Solution 1: Check the gain setting for the affected oxygen channel.
  • The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
  • The gain was correct: Locate the problem using the procedure described in Locating a Problem and Sensor test