Revision as of 16:50, 13 February 2015

Template:Technical service


Problems

  • The oxygen signal remains at or close to zero even at high oxygen levels.
  • The signal at air saturation is abnormally low.

Test and Solutions

The oxygen signal is very low at air saturation and less responsive than expected.

Solution 1: Check the gain setting for the affected oxygen channel.
  • The gain setting was too low, and the oxygen signal increases to the expected level after increasing the gain.
  • The gain was correct: Locate the problem using the procedure described in Locating a problem and Sensor test.