POS calibration: Difference between revisions

From Bioblast
No edit summary
(fill page)
Line 1: Line 1:
[[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]] Β 
DatLab converts the [[raw signal]] to oxygen concentrationsΒ  by applying conversion factors determined during a POS calibration comprising an [[air calibration]] and a [[zero calibration]]. For each [[Gain (O2 channel) | gain]] a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the [[Raw signal]].
Please see [[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]] Β 





Revision as of 12:07, 14 September 2010

DatLab converts the raw signal to oxygen concentrations by applying conversion factors determined during a POS calibration comprising an air calibration and a zero calibration. For each gain a different set of calibration values is necessary. Frequently, assumed hardware problems turn out to be a simple calibration problem. This can easily be detected by observing the Raw signal. Please see [MiPNet12.08]