:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? ''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? ''' | ||
::::::: After chamber reassembly you might experience a small leaky chamber only identified by a ''leaky chamber test''- »[ | ::::::: After chamber reassembly you might experience a small leaky chamber only identified by a ''leaky chamber test''- »[[O2k-Chamber_MitoPedia#Leaky_chamber_test|Leaky chamber test]]« | ||
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Blue-info.png|40px|Info: improve]] '''Do you change the [[Smoothing|slope smoothing]] to 20 while running experiments with biological sample?''' | :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Blue-info.png|40px|Info: improve]] '''Do you change the [[Smoothing|slope smoothing]] to 20 while running experiments with biological sample?''' |
Revision as of 10:48, 5 March 2020
- Do you perform the leaky chamber test after chamber assembly?
- After chamber reassembly you might experience a small leaky chamber only identified by a leaky chamber test- »Leaky chamber test«
- Do you perform the leaky chamber test after chamber assembly?
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - »Slope smoothing«
- Do you change the slope smoothing to 20 while running experiments with biological sample?
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
- By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with specific settings for the O2k-sV-Module. - »O2k-sV-Module«
- Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?