| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? '''
| |
| ::::::: After chamber reassembly you might experience a small leaky chamber only identified by a ''leaky chamber test''- ยป[https://www.bioblast.at/index.php/O2k-Chamber_MitoPedia#Leaky_chamber_test Leaky chamber test]ยซ
| |
|
| |
|
| :::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Blue-info.png|40px|Info: improve]] '''Do you change the [[Smoothing|slope smoothing]] to 20 while running experiments with biological sample?'''
| |
| ::::::: While running your experiment, especially during [[uncoupler titrations]], it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - ยป[[Smoothing#DatLab_oxygen_flux:_performance_and_data_analysis|Slope smoothing]]ยซ
| |