== In discussion for page ==


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how the "switch components" approach can guide you to find what is causing your O2k not to reach the required specifications?'''
::::::: Do not desperate if specifications given in the O2k-Manuals are not obtained. See how to perform an experimental troubleshooting on your own.-Β»[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]Β«
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?'''
::::::: See how to clean and dry the electrical connections of the OroboPOS-Connector.-Β»[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]Β«
== O2k-Open Support alert circulars ==
=== O2k-Open Support alert circular 2020(01): in prep ===
::'''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.'''
::::::: See how you can check that the two blue OroboPOS-Holders are intact. - Β»[[O2k-OroboPOS-Holder_MitoPedia#O2k-OroboPOS-Holder:_Leaky_chamber_due_to_damaged_OroboPOS-Holder |O2k-OroboPOS-Holder]]Β«
<br />
::'''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|40px|Take care]] '''Calibration parameters are transferred to experimental DatLab files'''
::::::: Save calibration information from selected [[Marks - DatLab |Marks]] while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - Β»[[Oxygen_calibration_-_DatLab#Oxygen_calibration_in_instrumental_and_experimental_DatLab_files | Oxygen calibration - DatLab]]Β«
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform O2k-Chamber volume calibration?'''
::::::: The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the [[O2k-sV-Module]]). - Β»[[O2k-Chamber MitoPedia]]Β«
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you use the correct O<sub>2</sub> solubility factor (F<sub>M</sub>) in the β€˜O2 calibration’ window in DatLab?
::::::: The O<sub>2</sub> solubility factor varies according to the medium used and influences the air calibration. - Β»[[Oxygen solubility factor]]Β«