Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:Setting the oxygen concentration"

From Bioblast
Line 1: Line 1:
{{Technical support}}
{{Technical support}}


The oxygen concentration can be increased or decreased during an experiment with biological samples, depending on the application (''e.g.,'' for achieving hypoxia, with permeabilized muscle fibers, etc) or simply whenever oxygen concentrations run low before the end of an experiment. There are different means to adjust oxygen concentrations:
The oxygen concentration can be increased or decreased during an instrumental or a biological experiment. The increase or decrease of oxygen concentration during an experiment performed in the presence of biological sample is relevant for specific applications (''e.g.'', hypoxia, tissue normoxia, high oxygen levels for working with permeabilized muscle fibers) or simply whenever an experiment run out of oxygen before is finished (re-oxygenations). There are different possibilities to adjust oxygen concentration in the Oroboros O2k:


=== Increasing the oxygen concentration ===  
=== Increasing the oxygen concentration ===  
It is always possible to increase the oxygen concentrations by combining catalase and injecting H<sub>2</sub>O<sub>2</sub> in the mitochondrial respiratory medium as described in  [[MiPNet14.13 Medium-MiR06| MiPNet14.3]]. By using the [[TIP2k]], the oxygen concentrations can be maintained between well defined limits, either using H<sub>2</sub>O<sub>2</sub> or, for very low oxygen concentrations, air saturated medium in the [[Microsyringe\500 mm3\TIP2k |TIP2k microsyringes]]. Such specification is called "oxystat" approach and supply appropriate templates for controlling the TIP2k. See also [[MiPNet12.10]].
*'''Increase oxygen levels by combining H<sub>2</sub>O<sub>2</sub> to a catalase containing medium (Medium-MiR06)'''


*'''Increase oxygen levels by adding H<sub>2</sub>O<sub>2</sub> to a catalase containing medium [[MiPNet14.13 Medium-MiR06|MiR06]]'''
:It is always possible to increase the oxygen concentrations by combining catalase and injecting H<sub>2</sub>O<sub>2</sub> in the mitochondrial respiration medium as described in  [[MiPNet14.13 Medium-MiR06| MiPNet14.3]]. By using the [[TIP2k]], the oxygen concentrations can be maintained between well defined limits, either using H<sub>2</sub>O<sub>2</sub> or, for very low oxygen concentrations, air saturated medium in the [[Microsyringe\500 mm3\TIP2k |TIP2k microsyringes]]. Such specification is called "oxystat" approach and supply appropriate templates for controlling the TIP2k. See also [[MiPNet12.10]].
:Add catalase to a final concentration of 280 IU/mL in the mitochondrial respiratory medium at the beginning of an experiment. When oxygen starts to become limited, inject 1-3 µL of [[Reoxygenation#Preparation_of_200_mM_H2O2_stock_solution|200 mM H<sub>2</sub>O<sub>2</sub> stock solution]]. The H<sub>2</sub>O<sub>2</sub> is quickly converted to O<sub>2</sub> with a high catalase concentration able to avoid any sample oxidative stress. During this procedure the O2k-Chamber is not opened and closed and decreasing disturbance of the system and allowing faster POS stabilization.
:Add catalase to a final concentration of 280 IU/mL in the mitochondrial respiratory medium at the beginning of an experiment. When oxygen starts to become limited, inject 1-3 µL of [[Reoxygenation#Preparation_of_200_mM_H2O2_stock_solution|200 mM H<sub>2</sub>O<sub>2</sub> stock solution]]. The H<sub>2</sub>O<sub>2</sub> is quickly converted to O<sub>2</sub> with a high catalase concentration able to avoid any sample oxidative stress. During this procedure the O2k-Chamber is not opened and closed and decreasing disturbance of the system and allowing faster POS stabilization.
 
:The initial increase in oxygen, however, is preferentially made by injecting oxygen, since there is the risk of bubble formation if the oxygen concentration is increased in a single large step. If oxygen gas is not available for the initial oxygenation, a very small bubble may be left in the chamber while slowly rising the oxygen level to 500 µM with additions of H<sub>2</sub>O<sub>2</sub>, such that gas can escape into the small bubble and then be extruded by fully closing the chamber. During the experiment re-oxygenations must be progressively performed in order to avoid gas bubble formation.
The initial increase in oxygen, however, is preferentially made by injecting oxygen, since there is the risk of bubble formation if the oxygen concentration is increased in a single large step. If oxygen gas is not available for the initial oxygenation, a very small bubble may be left in the chamber while slowly rising the oxygen level to 500 µM with additions of H<sub>2</sub>O<sub>2</sub>, such that gas can escape into the small bubble and then be extruded by fully closing the chamber. During the experiment re-oxygenations must be progressively performed in order to avoid gas bubble formation.


* '''Increase oxygen levels with injection of oxygen into the gas phase'''  
* '''Increase oxygen levels with injection of oxygen into the gas phase'''  
If you replace the gas phase above the liquid phase by pure oxygen you can increase oxygen levels above air saturation, as it is recommended for measuring mitochondrial respiratory function in muscle biopsies.
:If you replace the gas phase above the liquid phase by pure oxygen you can increase oxygen levels above air saturation, as it is recommended for measuring mitochondrial respiratory function in muscle biopsies.
:* [[Open chamber]]: Before injecting oxygen, the O2k-Chamber needs to be opened to obtain a defined gas phase above the aqueous phase.  
:* [[Open chamber]]: Before injecting oxygen, the O2k-Chamber needs to be opened to obtain a defined gas phase above the aqueous phase.  
:* Using the[[Syringe\60 ml\Gas-Injection | 60 mL syringe]] oxygen is injected into the gas phase of the O2k-Chamber.
:* Using the[[Syringe\60 ml\Gas-Injection | 60 mL syringe]] oxygen is injected into the gas phase of the O2k-Chamber.
Line 18: Line 17:


*'''Increase oxygen levels by [[Open_chamber | opening the chamber]]'''
*'''Increase oxygen levels by [[Open_chamber | opening the chamber]]'''
 
:Lift the stopper using the [[Stopper-Spacer | Stopper-Spacer tool]] to a standard position with a fixed gas phase above the aqueous phase in the O2k-Chamber. Leave the O2k-Chamber open till oxygen concentration reaches an approximate level of the air calibration. Close the O2k-Chamber by closing the stoppers completely and wait till the POS is stable again (~ 5-10 minutes).  
:Lift the stopper using the [[Stopper-Spacer | Stopper-Spacer tool]] to a standard position with a fixed gas phase above the aqueous phase in the O2k-Chamber. Leave the O2k-Chamber open till oxygen concentration reaches an approximate level of the air calibration. Close the O2k-Chamber by closing the stoppers completely and wait till the POS is stable again (~ 5 - 10 minutes).  
:If possible, it is preferable to re-oxygenate in a phase of low measured respiratory activity since minor amounts of oxygen are consumed during the stabilization phase of the POS after closing the O2k-Chamber.  
:If possible, it is preferable to re-oxygenate in a phase of low measured respiratory activity since minor amounts of oxygen are consumed during the stabilization phase of the POS after closing the O2k-Chamber.  




=== Decreasing the oxygen concentration ===  
=== Decreasing the oxygen concentration ===  
Low oxygen levels can be reached depending on the application. During an experiment with a biological sample this may be performed by injection of N<sub>2</sub> in the gas phase, followed by closing the stoppers once, as described below:
:Low oxygen levels can be reached depending on the application. During an experiment with a biological sample this may be performed by injection of nitrogen (N<sub>2</sub>) in the gas phase, followed by closing the stoppers once the desire oxygen level was reached, as described below:
:* [[Open chamber]]: Before injecting nitrogen, the O2k-Chamber needs to be opened to obtain a defined gas volume above the aqueous phase.  
:* [[Open chamber]]: Before injecting nitrogen, the O2k-Chamber needs to be opened to obtain a defined gas volume above the aqueous phase.  
:* Using the [[Syringe\60 ml\Gas-Injection | 60 mL syringe]] nitrogen is injected into the gas phase of the O2k-Chamber.
:* Using the [[Syringe\60 ml\Gas-Injection | 60 mL syringe]] N<sub>2</sub> is injected into the gas phase of the O2k-Chamber.
:* Close the chamber when the O<sub>2</sub> concentration approaches the desired O<sub>2</sub> level.  
:* Close the chamber when the O<sub>2</sub> concentration approaches the desired O<sub>2</sub> level.  
However, often the desired starting value is simply reached by waiting until the sample has consumed the required amount of oxygen, when the "oxystat" follows.
:However, often the desired starting value is simply reached by waiting until the sample has consumed the required amount of oxygen, when the "oxystat" follows.


For calibration and for  [[MiPNet14.06 Instrumental O2 background |instrumental O<sub>2</sub> background tests]] purposes, dithionite is used to reduce oxygen levels, however this is not recommended in the presence of biological samples.
:To perform [[Zero calibration|zero calibration]] oxygen should be removed by the addition of [[Dithionite|dithionite]] or biological sample.
:To reduce the oxygen levels during an [[MiPNet14.06 Instrumental O2 background |instrumental O<sub>2</sub> background test]], [[Dithionite|dithionite]] should be used.

Revision as of 11:04, 9 June 2020


                  


O2k-Open Support

Talk:Setting the oxygen concentration



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



The oxygen concentration can be increased or decreased during an instrumental or a biological experiment. The increase or decrease of oxygen concentration during an experiment performed in the presence of biological sample is relevant for specific applications (e.g., hypoxia, tissue normoxia, high oxygen levels for working with permeabilized muscle fibers) or simply whenever an experiment run out of oxygen before is finished (re-oxygenations). There are different possibilities to adjust oxygen concentration in the Oroboros O2k:

Increasing the oxygen concentration

  • Increase oxygen levels by adding H2O2 to a catalase containing medium MiR06
It is always possible to increase the oxygen concentrations by combining catalase and injecting H2O2 in the mitochondrial respiration medium as described in MiPNet14.3. By using the TIP2k, the oxygen concentrations can be maintained between well defined limits, either using H2O2 or, for very low oxygen concentrations, air saturated medium in the TIP2k microsyringes. Such specification is called "oxystat" approach and supply appropriate templates for controlling the TIP2k. See also MiPNet12.10.
Add catalase to a final concentration of 280 IU/mL in the mitochondrial respiratory medium at the beginning of an experiment. When oxygen starts to become limited, inject 1-3 µL of 200 mM H2O2 stock solution. The H2O2 is quickly converted to O2 with a high catalase concentration able to avoid any sample oxidative stress. During this procedure the O2k-Chamber is not opened and closed and decreasing disturbance of the system and allowing faster POS stabilization.
The initial increase in oxygen, however, is preferentially made by injecting oxygen, since there is the risk of bubble formation if the oxygen concentration is increased in a single large step. If oxygen gas is not available for the initial oxygenation, a very small bubble may be left in the chamber while slowly rising the oxygen level to 500 µM with additions of H2O2, such that gas can escape into the small bubble and then be extruded by fully closing the chamber. During the experiment re-oxygenations must be progressively performed in order to avoid gas bubble formation.
  • Increase oxygen levels with injection of oxygen into the gas phase
If you replace the gas phase above the liquid phase by pure oxygen you can increase oxygen levels above air saturation, as it is recommended for measuring mitochondrial respiratory function in muscle biopsies.
  • Open chamber: Before injecting oxygen, the O2k-Chamber needs to be opened to obtain a defined gas phase above the aqueous phase.
  • Using the 60 mL syringe oxygen is injected into the gas phase of the O2k-Chamber.
  • Close the chamber when the O2 concentration approaches the desired O2 level.
Lift the stopper using the Stopper-Spacer tool to a standard position with a fixed gas phase above the aqueous phase in the O2k-Chamber. Leave the O2k-Chamber open till oxygen concentration reaches an approximate level of the air calibration. Close the O2k-Chamber by closing the stoppers completely and wait till the POS is stable again (~ 5-10 minutes).
If possible, it is preferable to re-oxygenate in a phase of low measured respiratory activity since minor amounts of oxygen are consumed during the stabilization phase of the POS after closing the O2k-Chamber.


Decreasing the oxygen concentration

Low oxygen levels can be reached depending on the application. During an experiment with a biological sample this may be performed by injection of nitrogen (N2) in the gas phase, followed by closing the stoppers once the desire oxygen level was reached, as described below:
  • Open chamber: Before injecting nitrogen, the O2k-Chamber needs to be opened to obtain a defined gas volume above the aqueous phase.
  • Using the 60 mL syringe N2 is injected into the gas phase of the O2k-Chamber.
  • Close the chamber when the O2 concentration approaches the desired O2 level.
However, often the desired starting value is simply reached by waiting until the sample has consumed the required amount of oxygen, when the "oxystat" follows.
To perform zero calibration oxygen should be removed by the addition of dithionite or biological sample.
To reduce the oxygen levels during an instrumental O2 background test, dithionite should be used.