Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

O2k-Titration Set

From Bioblast
Revision as of 09:08, 12 February 2013 by Eigentler Andrea (talk | contribs)

Template:Mainpage orobopedia

                



O2k-Titration Set

O2k-Catalogue

Description The Titration Set consists of 6x 10 µl, 6x 25 µl, and 2x 50 µl Hamilton glass syringes, and 3 spare plungers for the 10 µl syringes. It is a basic HRR-Accessory included in the O2k-Core Unit.
Product ID 20820-01
Type O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue
Link MiPNet09.12 O2k-Titrations
Image
Hamilton Syringes for Manual Titration.jpg


The Titration Set consists of

6 Microsyringe\10 mm3 51/0.13 mm 10 mm3
6 Microsyringe\25 mm3 51/0.15 mm 25 mm3
2 Microsyringe\50 mm3 51/0.15 mm 50 mm3
3 Plunger\10 mm3 for 10 mm3 syringe


Treatment of the syringes

  • for long time storage of the syringes fill them with 100 % ethanol and store them in 15 ml Falcons filled with 100 % ethanol that the needle and the beginning of the glass syringe are covered with liquid
  • before using the syringes always rinse them outside with the solvent (water or ethanol) of the chemical
  • before using the syringes wash them 3 times with the solvent - be aware that the whole syringe is filled till the end of the plunger!
  • between titration: rinse the outside with water (to get rid of the respiration medium) and if necessary with ethanol depending on the solvent of the chemical

•Separate inhibitors from substrates •Between two runs during the day: rinse outside with water •End of the day: 3 x solvent, 3 x EtOH 100% •storage: in 100 % ethanol