Difference between revisions of "Run DL-Protocol/Set O2 limit"

From Bioblast
m
m
Line 22: Line 22:
::::'''Set DL-Protocol / O2 limit'''. In the 'Set DL-Protocol / O2 limit' window a DL-Protocol can be assigned to specific chamber.
::::'''Set DL-Protocol / O2 limit'''. In the 'Set DL-Protocol / O2 limit' window a DL-Protocol can be assigned to specific chamber.


:::::*'''Buttons'''. DL-protocols (either a .DLP or .DLPU file) can be selected ([[File:Select-button.png|50px]]) for chamber A and B. '.DLP' is the original file formate for DL-Protocols provided by Orobors, '.DLPU' files are user-modified versions (see [[#DL-Protocols-anchor3|A: Export DL-Protocol]] and [[#DL-Protocols-anchor2|'Edit DL-Protocol' window]]) of a specific DL-Protocols ('.DLP' file). Only when a DL-Protocol is initially selected, the [[File:Show_this_file-button.png|60px]] button ist active and enables to automatically open the DL-Protocol file in another DatLab instance to see a DL-Protocol specific example data plot with properly set marks and events. Clicking the [[File:Clear-button.png|50px]] button removes the selected DL-Protocol.
:::::*'''Buttons'''. DL-protocols (either a .DLP or .DLPU file) can be selected ([[File:Select-button.png|50px]]) for chamber A and B. '.DLP' is the original file format for DL-Protocols provided by Orobors, '.DLPU' files are user-modified versions (see [[#DL-Protocols-anchor3|A: Export DL-Protocol]] and [[#DL-Protocols-anchor2|'Edit DL-Protocol' window]]) of a specific DL-Protocols ('.DLP' file). Only when a DL-Protocol is initially selected, the [[File:Show_this_file-button.png|60px]] button ist active and enables to automatically open the DL-Protocol file in another DatLab instance to see a DL-Protocol specific example data plot with properly set marks and events. Clicking the [[File:Clear-button.png|50px]] button removes the selected DL-Protocol.


:::::*'''Additional information'''. When a DL-protocol is selected for chamber A or B further text information is provided in the chamber specific field: (1) protocol name, (2) a hyperlink (in blue, underlined) to the protocol in the Bioblast and (3) additional information and short description for the DL-Protocol.
:::::*'''Additional information'''. When a DL-protocol is selected for chamber A or B further text information is provided in the chamber specific field: (1) protocol name, (2) a hyperlink (in blue, underlined) to the protocol in the Bioblast and (3) additional information and short description for the DL-Protocol.

Revision as of 16:02, 26 July 2017


high-resolution terminology - matching measurements at high-resolution


Run DL-Protocol/Set O2 limit

Description

DL-Protocols can be selected in DatLab 7 in the pull-down menu 'Protocol': Set DL-Protocol / O2 limit. A DL-Protocol defines the sequence of Events and Marks. Instrumental DL-Protocols are used for calibrations and instrumental quality control, typically without experimental sample in the incubation medium. DL-Protocols for substrate-uncoupler-inhibitor titrations (see MitoPedia: SUIT) proceed stepwise to activate a sequence of coupling control states and pathway control states. A specific SUIT protocol can be assigned to O2k-chamber A or B or both. The Titration-Injection-microPump TIP2k can be programmed for automatic control of titration steps in a DL-Protocol. A collection of evaluated and tested standard DL-Protocols are provided. These can be edited and saved under 'Lab protocols' or under the User logged into DatLab. A Lower O2 limit [µM] can be defined for each chamber, to trigger an automatic warning when the experimental O2 concentration declines below this limit as a WARNING to remind the user that re-oxygenation of the chamber is required.

Abbreviation: DLP





MitoPedia O2k and high-resolution respirometry: DatLab 



Protocol menu

Show DL-Protocol can be check-marked (Check-mark.png) to hide or show the DL-Protocol window. Alternatively, the Button show-hide DL-Protocol window.png button (upper left) in the DL-Protocol window can be used to hide or show the window. When DatLab is started for data recording, an empty DL-Protocol window shows up because no DL-protocol is specified at this moment.


Synchronous DL-Protocol events. This option is only available when the same DL-Protocol is selected (in 'Set DL-Protocol / O2 limit') for chamber A and B (protocol names are displayed in the name field of the DL-Protocol window). When checked Check-mark.png, the user only needs to follow one protocol sequence of either chamber A or B. For example, when an event is set for chamber A, it will also be automatically set for chamber B - the 'Event Information' window indicates "Set event in both chambers" in bold letters.
Set DL-Protocol / O2 limit For chamber A (upper frame) a specific protocol is selected, showing the name ("RP1mt"), the hyperlink to the biobast in blue and additional information. Furthermore, the 'Lower O2 limit [µM]' alert is checked. In chamber B (lower frame) no DL-protocol is selected.


Set DL-Protocol / O2 limit. In the 'Set DL-Protocol / O2 limit' window a DL-Protocol can be assigned to specific chamber.
  • Buttons. DL-protocols (either a .DLP or .DLPU file) can be selected (Select-button.png) for chamber A and B. '.DLP' is the original file format for DL-Protocols provided by Orobors, '.DLPU' files are user-modified versions (see A: Export DL-Protocol and 'Edit DL-Protocol' window) of a specific DL-Protocols ('.DLP' file). Only when a DL-Protocol is initially selected, the Show this file-button.png button ist active and enables to automatically open the DL-Protocol file in another DatLab instance to see a DL-Protocol specific example data plot with properly set marks and events. Clicking the Clear-button.png button removes the selected DL-Protocol.
  • Additional information. When a DL-protocol is selected for chamber A or B further text information is provided in the chamber specific field: (1) protocol name, (2) a hyperlink (in blue, underlined) to the protocol in the Bioblast and (3) additional information and short description for the DL-Protocol.
O2-limit warning button
  • Lower O2 limit [µM]. When checked, an automatic warning notifies the user when the actual oxygne leves falls below the specified value for the accoring chamber: A red flashing "WARNING" button in the status line (DatLab window bottom, right) appears and a "WARNING" event is set in the plot. To reset the actual warning status (to notify the next possible fall-below), the red flashing "WARNING" button has to be clicked. First, a 'Warning window' opens to show all warnings during the experiment. After closing the window, the reset ist completed.


A: Export DL-Protocol. The user can save changes in a DL-Protocol and export an user-modified DL-Protocol (.DLPU) for reuse. To make changes in a DL-Protocol see: 'Edit DL-Protocol' window
B: Export DL-Protocol. applies for chamber B as described above in 'A: Export DL-Protocol'.


DL-Protocol window

DL-Protocol window in DatLab, highlighted by a red box in the image. The DL-Protocol windows for chamber A (up) and chamber B (down) are shown with different DL-protocols as indicated by the DL-protocol name in the name field (1, red oval)
The DL-Protocol window appears to the very right for every chamber. It consisits of three buttons at the top, a name field and a list to display events and marks.
  • Buttons. The Button show-hide DL-Protocol window.png button is used to hide/show the DL-Protocol window (similar to 'Show DL-Protocol' in the 'Protocol' menu). Clicking the Hide marks.png/Show marks.png button toggles between hiding or showing the marks in the list. When the marks in the DL-Protocol are hidden the user only needs to focus on the sequence of events, facilitating the experimental procedure.The Edit.png button opens the 'Edit DL-Protocol' window to make changes in events and marks in the DL-Protocol.
  • Information. Below the buttons, the 'name field' automatically displays the name of the loaded protocol.
  • List. The list in the DL-Protocol window shows the sequence of events and marks. The numbers in the event and mark names corresponds to the defined sequence in the DL-Protocol. Mark names are preceded by ">> M:", events are indicated by ">> E:" or " E:". Every mark or event is unique and must be set only once. However, there is an exception for events identified by "", which can be set consecutively multiple times (multi-events), as it is necessary for titrating the uncoupler (U) in SUIT protocols.
Alt text
Edit DL-protocol window ...
'Edit DL-Protocol' window. In the listed events and marks, changes can be made in 'Final conc. [mM]', 'Titration vol. [µL]' and 'Multi' but not in 'Type', 'Name' and 'Definition/State'. Further changes are accepted in 'Comment' and 'DL-Protocol description' at the bottom of the window.
  • Information in the window head. The active plot is indicated in the according name field (window, top). A second name field displays the DL-Protocol name, furthermore, a hyperlink (blue, underlined) to the biobast is avialable.
  • List. In the list of the window, events are always displayed, however, only the marks (if available) for the active plot (O2 or O2 flux) are shown.
    • Type and Name. Similar to the DL-Protocol window, mark names are preceded by ">> M:", events are indicated by ">> E:" or " E:" (multi-events).
    • Final conc. [mM] and Titration vol. [µL]. The fields contain the according values for an event or mark and can be changed by the user.
    • Multi. To toggle an events status from or to the status 'multi-event' ,the checkbox in the column has to be set List-ckeckbox-checked.png or unchecked List-ckeckbox-unchecked.png, wherupon the indicator (">>" or " ") for the event changes. For marks, a set checkbox is of no relevance.
  • Comment. This field contains specific information for an event or mark and can be changed/edited by the user.
  • DL-Protocol description. This field contains general, DL-Protocol specific information and can be changed/edited by the user.
  • Changes can be accepted Save-button.png or rejected Chancel-button.png. Finally, changes in the DL-Protocol can be saved/exported to generate a user modified DL-Protocol (.DLPU), see A: Export DL-Protocol.


Running a DL-Protocol

  • Start. Hardware and Software must be setup properly (refer to the O2k-FluoRespirometer manual). DatLab 7 has to be started, connected ('Connect to O2k') and all instrumental- and experimental parameters have to be set.
  • DL-Protocol basics.
    • The 'DL-Protocol window' shows the seqence of DL-Protocol events and DL-Protocol marks. For a better understanding, when running a DL-Protocol it is possible to set user-defined marks and user-defined events (menu 'Experiment' -> 'Add event' or F4 or STRG-left-click in plot), which will not affect the DL-Protocol but will be stored togehter with the DL-Protocol marks and events in the .DLD data file. It is possible to change a user-defined mark or user-defined event into DL-Protocol mark or DL-Protocol event, respectively.
    • An HRFR experiment is primarily an event-driven process, thus, DL-Protocol marks don't need to be regarded at first, when running the DL-Protocol. Hence, for practicability, the DL-Protocol marks can be hidden in the list ('DL-Protocol window': Hide marks.png) and the user only has to focus on the sequence of the DL-Protocol events. However, DL-Protocol marks or any user-defined marks can be set at any time during the experimental run.
  • DL-Protocol colors. In the process of running a DL-Protocol following the DL Protocol events (>> E: or  E:) and even after, when setting DL Protocol marks (>> M:), events and marks appear in different colors according to the following rules:


    • Aqua:
    • Light green:
    • Red:
    • White:
Color type indicator Description
Yellow events ">> E:" or " E:" indicates the next DL-Protocol event going to be completed
Text event ">> E:" DL-Protocol event is completed and the position of the actual event or defined order of all events in the DL Protocol has not been violated (e.g. wrong position, any missing events in the processed seuqence, ...)
Aqua mulit event " E:" DL-Protocol multi event is completed and position of the actual multi event or defined order of all events in the DL Protocol has not been violated (e.g. wrong position, any missing events in the processed seuqence, ...). As intended, the multi event (in auqua) can be set consecutively multiple times before processing with the next DL-Protocol event (in yellow) in the DL-Protocol
Light green mark ">> M:" zzz
Red marks, events ">> M:", ">> E:" or " E:" zzz
White marks, events ">> M:", ">> E:" or " E:" zzz










  • Running a DL-Protocol. Once the DL-Protocol is loaded, the protocol layout will be displayed on the right side in the 'DL-Protocol window'. The experiment can now be started following the DL Protocol event sequence. For this purpose, left-klick onto the first DL Protocol event (in yellow) of the DL Protocol and the Event information window will open.


Titrations in SUIT protocols

  • Stock concentration of substance B in the titration syringe, cB,in [mM] or [µM] as specified.
  • Titration volume, v [µl]
  • Chamber volume, V [µl]
  • Final concentration, cB [mM] or [µM] as specified.
cB [mM] = cB,in [mM] x v [µl] / V [µl]
» MiPNet09.12 O2k-Titrations
Example: Malate (M), cM,in = 800 mM stock; v = 5 µl titration volume; V = 2,000 µl chamber volume.
cM [mM] = cM,in [mM] x v [µl] / V [µl] = 800 mM x 5 µl / 2,000 µl = 2 mM