Difference between revisions of "Stirrer-Bar\white PVDF\15x6 mm"
From Bioblast
m (Protected "Stirrer-Bar\white PVDF\15x6 mm" ([edit=sysop] (indefinite) [move=sysop] (indefinite))) |
|||
Line 5: | Line 5: | ||
|product id=23210-01 | |product id=23210-01 | ||
|product type=O2k, O2k-Assembly Kit | |product type=O2k, O2k-Assembly Kit | ||
|info= | |info=[[MiPNet06.01 O2k-Overview]] | ||
|product image=[[Image:Stirrer bar PVDF.jpg|right|180px|link=http://www.bioblast.at/index.php/Stirrer-Bar%5Cwhite_PVDF%5C15x6_mm]] | |product image=[[Image:Stirrer bar PVDF.jpg|right|180px|link=http://www.bioblast.at/index.php/Stirrer-Bar%5Cwhite_PVDF%5C15x6_mm]] | ||
}} | }} |
Revision as of 16:39, 11 March 2014
Stirrer-Bar\white PVDF\15x6 mm
Description | Stirrer-Bar\white PVDF\15x6 mm, operated in the O2k-Chamber at constant stirring speed (standard is 750 rpm, or 12.5 Hz), to provide optimum mixing of the sample in the aqueous medium and ensure a stable signal of the polarographic oxygen sensor (OroboPOS) placed in a position of maximum current of the medium.
O2k-Core: 2 in O2k-Assembly Kit. |
---|---|
Product ID | 23210-01 |
Type | O2k, O2k-Assembly Kit |
Link | MiPNet06.01 O2k-Overview |
Image |
- For dispatch of the O2k-Core, the Stirrer-Bars are shipped in the O2k-Service Box.
A "jumping" stirring bar problem is frequently caused by small particles sticking to the glass chamber, not the stirring bar itself, see Cleaning the glass chamber.